At the completion of the device test, a
TEST FAILED
indicates a problem that should be
taken care of before the device is released for a night study.
DEVICE TEST 22:50
ID Error
battery=low
pat=bad led
pat=bad photo
file=unloaded
<-Back ->More
TEST FAILED 1:54
The following are the possible error, warning or information messages:
•
File error: not loaded, missing – the study file was not loaded or somehow
the file was deleted
•
File error: used x/3 x=1..3 – only when multi-night option is selected
•
Battery error: low – needs replacement of battery
•
Probe error: used, missing, bad – connect an unused probe
•
Hardware (H/W) error: error code - contact customer support
•
SBP/RESBP (Snore and Body Position sensor) warning: sensor missing –
does not affect PASSED status
•
RTC (Real Time Clock) error: faulty – indicates problem with internal clock
and need to set the clock (through the software)
•
Bracelet error: missing – the study file was chosen with the bracelet option
but the bracelet is not connected during the device test
•
Information messages:
o
multi-night=on - when a multi night study is required
o
bracelet=on - when a study with tamper-proof patient
identification bracelet is required
More->
indicates that there are more error/warning messages and will be displayed if the
Right (
->
) button is pressed.
<-Back
will move to the previous screen if the Left (
<-
) button is pressed.