Chapter 3 TEST MENU
3-18
IL-EMZ Service Manual
Adjustment in threshold value detection mode
1. Touch the DPM button by a test mode and see the
value of a LABEL SENSOR DPM.
TEST MENU [PRINTER ADJUSTMENT]
2. Operate the LABEL SPECIFICATION 2/2 of System
Setting menu.
Touch a button in the following procedure.
[System Setting menu] -> [LABEL SPEC.] ->
[9][9][9] -> [PAGE]
3. Enter the value of [LABEL SENSOR DPM] at numeric
keys and touch the [Label sensor DPM] field.
4. Enter the temporary threshold value 120 at numeric
keys and touch the [Label sensor DPM] field.
5. Press the [FEED] key 2 or 3 times. Then, label comes
out.
Enter the 2000 at numeric keys and touch the [X] key.
Then, graph pops up.
FEED
Label gap
2
0
0
0
X
Label gap