Chapter 3 TEST MENU
3-16
IL-EMZ Service Manual
1. The adjustment method of a gain.
Touch the [LABEL SENSOR LEVEL] button.
Display field will change to green.
Touch the (
UV
) adjustment buttons. Touch this button
again to change the display to white and set adjustment
data.
2. T
he check method of a sensor level.
Touch the [FEED] key 2 or 3 times. Then, label comes
out.
Enter the 2000 at numeric keys and press the [X] key.
Then, graph pops up.
FEED
Label gap
2
0
0
0
X
Label gap