Read out the measurement values
Read measured output voltage:
:MEASure:VOLTage?
UM,
RANGE=3.000kV,
VALUE=2.458kV
Read measured output current:
:MEASure:CURRent?
IM,
RANGE=100.0mA,
VALUE=25.3mA
Read out the device status
:READ:STATus?
DI,
b
15
b
14
b
13
b
12
b
11
b
10
b
9
b
8
b
7
b
6
b
5
b
4
b
3
b
2
b
1
b
0
For status field description, see chapter 6.1, device status.
Read out the LAM status
:READ:LAM?
For LAM status description, see chapter 6.1, LAM status.
Read out the device identification
:READ:IDNT?
ID,
iseg Spezialelektronik
r5.01
sn.680042
Typ
HPn
30
107
6.4 Common command set (Compatibility to old HPS)
The Common commands can be used all the time, independently from selected instruction set (ET or
SCPI).
A response example for commands that returns an answer is shown at the right side (
printed Italic
).
Command
Response example:
Read out the device identification
*IDN?
ID,
iseg
Spezialelektronik
r5.01
sn.680043
Typ
HPn
30
107
Device control
*LLO
Local Lockout: disable local buttons
*GTL
Go to Local: enable local buttons
*CLS
Clear Status: clear error status
*RST
Reset: Device is rebooting
Read/change Instruction set
*INSTR?
Query current instruction set
Instruction type,
SCPI
*INSTR,SCPI
Change to SCPI instruction set
*INSTR,ET
Change to ET instruction set
Read/change RS-232 Echo state
*ECHO?
Query Echo status
Echo
off
*ECHO*ON
Turn on Echo
*ECHO*OFF
Turn off Echo