6.3 SCPI command set (Compatibility to old HPS)
This instruction set ist selected with Menu “F10 Set Instruct“ “SCPI“.
Setting commands
Command:
C
ommand example:
Command to set an output voltage:
:VOLTage <Voltage>kV
:VOLT
1.000kV
Command to set a voltage limit:
:LIMIT:VOLTage <Voltage>kV
:LIMIT:VOLT
1.000kV
Command to set an output current:
:CURRent <Current>mA
:CURR
30mA
Command to set a current limit:
:LIMIT:CURRent <Current>mA
:LIMIT:CURR
30mA
Command to set a voltage ramp speed:
:CONFigure:RAMP <Rampspeed>V/s
:CONF:RAMP
3000V/s
Command to turn high voltage on and off:
:VOLTage ON
:VOLTage OFF
Command for Emergency off (The HV generation is shut off permanently and the set values for volt-
age and current are set to Zero):
:VOLTage EMCY OFF
Command for setting KILL function “Enable“ or “Disable“:
:CONFigure:KILL ENable
:CONF:KILL
EN
:CONFigure:KILL DISable
Read out the setting commands
Command:
Response example:
Command to read out the set voltage:
:READ:VOLTage?
U,
RANGE=3.000kV,
VALUE=2.458kV
Command to read out the voltage limit:
:READ:LIMIT:VOLTage?
UL,
RANGE=3.000kV,
VALUE=2.458kV
Command to read out the set current:
:READ:CURRent?
I,
RANGE=100.0mA,
VALUE=30.0mA
Command to read out the current limit::
:READ:LIMIT:CURRent?
IL,
RANGE=100.0mA,
VALUE=30.0mA
Command to read out the voltage ramp speed:
:READ:RAMP?
Ramp, RANGE=3000V/s,
VALUE=3000V/s