Problem
Action
No treatment beam
• Ensure that the Fiber Switch
is in the correct position for
the console and wavelength
you are using.
• Ensure that the remote
interlock has not been
activated.
• Ensure that the aiming beam
is present and bright.
• If using the OcuLight GLx or
TX console, ensure that the
eye safety filter is in the
closed position.
• If you still have no treatment
beam, contact your local
IRIDEX Technical Support
representative.
The Status panel reads:
Connect Fiber
• Ensure that the
OcuLight Symphony Adapter
is properly connected.
Connect SmartKey
• Ensure that the SmartKey is
properly inserted.
Spot Size?
• Ensure that the spot size
selector is not between
detents.
Green Safety Filter?
• Ensure that the eye safety filter
is properly connected and, if
using the OcuLight GLx or TX
console, that it is in the closed
position.
SWITCH XXXXX Fiber on SLA
Press Mode to Continue
• Ensure that the Fiber Switch
is in the correct position for
the console and wavelength
you are using and press the
Mode button on the
OcuLight SLx/GLx/TX
console to continue.