29
Operation
31144C-EN
If desired, attach the finger-rest to the slit lamp
Slide the finger-rest onto the binocular support post and tighten the
thumbscrew to secure.
0
15
30 45 60 75 90
105
120
12
8
20
5
12
32
Finger-rest
Binocular support
post
Thumbscrew
WARNING
Always inspect the
fiber-optic assembly before
securing it to the slit lamp
headrest to ensure that it
has not been damaged. A
damaged fiber-optic cable
could cause accidental laser
exposure or injury to
yourself, your patient, or
others in the treatment
room.
NOTE
Avoid making any sharp
bends or kinks in the cable
assembly as this can
compromise the integrity of
the fiber-optic and prevent
or reduce laser light
transmission.
Secure the fiber-optic assembly
Secure the fiber-optic assembly to the slit lamp
Using the fiber-optic fasteners, secure the fiber-optic assembly to the slit
lamp headrest by forming a gentle arc with no less than a 15 cm (6 inch)
diameter bend.
Fiber-optic fastener
Fiber-optic cable
810 nm
532 n
m
810-LS
Fib
er Sw
itch
15 cm