25
Operation
31144C-EN
REFERENCE
See your CSO slit lamp
manual for instructions on
assembling the slit lamp.
Installing the OcuLight Symphony Adapter
Prepare the slit lamp for OcuLight Symphony Adapter installation
1. Ensure that the slit lamp is properly assembled.
2. Loosen the thumbscrew to remove the oculars from the slit lamp.
3. Rotate the illumination tower to either side of the centered position,
allowing enough clearance for the IRIDEX split-mirror illumination
prism, if applicable, and the OcuLight Symphony Adapter.
120
105
90
75
60
45
30
15
0
12
8
20
5
12
32
Install the IRIDEX split-mirror illumination prism
(for Zeiss 30 SL and 30 SL/M slit lamps only)
Replace the original illumination prism with the IRIDEX split-mirror
illumination prism:
1. Loosen the retaining ring at the base of the original illumination prism by
rotating the knurled ring clockwise until it disengages.
2. Lift the original prism out of the illumination housing and properly store.
3. Lower the IRIDEX split-mirror illumination prism into the housing,
aligning the pin at the base of the prism with the slot in the housing.
4. Rotate the knurled ring counter-clockwise to secure the IRIDEX split-
mirror illumination prism in position.
IRIDEX split-mirror illumination prism
Zeiss 30 SL model shown
Illumination housing
Summary of Contents for OcuLight Symphony
Page 1: ...OcuLight Symphony IRIDEX 31144C EN Operator Manual...
Page 6: ...iv OcuLight Symphony Operator Manual 31144C EN...
Page 12: ...6 OcuLight Symphony Operator Manual 31144C EN...
Page 56: ...50 OcuLight Symphony Operator Manual 31144C EN...
Page 68: ...62 OcuLight Symphony Operator Manual 31144C EN...