TIGER Select Instrument User Manual V3.2
Unrivalled Gas Detect
ion.
Page 61 of 73
ionscience.com
1.
Ensure TIGER Select is switched Off.
2.
Remove the Instrument Boot (see
3.
Remove the AA Battery Pack.
4.
Remove the screws retaining the battery cover
and lift it off. This reveals the batteries in the
Battery Holder.
5.
Remove the Battery Holder.
6.
Replace the batteries. Use Industrial by Duracell
ID1500 batteries.
7.
Check that the polarity of the batteries is correct.
Replace the Battery Holder.
8.
Refit the battery cover and the securing screws.
9.
Replace the Instrument Boot ensuring that it is
snug around the front of the Instrument Body
and the lights are not obscured.
WEEE REGULATIONS
Dispose of used batteries in accordance with all local and national safety and
environmental requirements.
Inlet Probe Assembly
All parts of the probe assembly can be replaced if they become damaged or contaminated with use.
1
Inlet Probe
2
Filter Clamp
3
Filter Housing Cap
4
O-ring
5
Probe Seal
6
PTFE Filter Disc
PTFE Filter Disc
It is essential that the TIGER Select is always used with a supplied 0.5 micron
PTFE Filter Disc fitted to the front of the instrument. Without the filter, particles
of debris and dust can be drawn into the MiniPID 2 Sensor inhibiting the
function of the instrument. These filters are consumable and should be changed
after every 100 hours of use. The frequency of replacement should be increased
for dusty or moisture laden environments. PTFE Filter Discs are available from
your distributor or at