TIGER Select Instrument User Manual V3.2
Unrivalled Gas Detect
ion.
Page 55 of 73
ionscience.com
5.
SPAN 1 calibration
: Press
Enter
. The gas and concentration for SPAN 1 (previously
set up in TigerPC) are displayed along with a 30 second countdown. Attach the SPAN 1 gas
using the supplied Calibration Adaptor and press
Enter
to start the SPAN 1 countdown.
6.
At the end of the countdown a tick '
✓
' will appear, indicating that the SPAN 1 has been
accepted. Press
Enter
. For a two-point calibration, the calibration procedure is complete.
7.
SPAN 2 calibration
: For a three-point calibration, the gas and concentration for
SPAN 2 are displayed (previously set up in TigerPC) along with a 30 second countdown.
Attach the SPAN 2 gas and press Enter to start the SPAN 2 countdown.
To skip SPAN 2, press
Esc
and then ‘skip’ to complete a SPAN 1 only calibration and return
to the main running screen.
8.
At the end of the count down a tick '
✓
' will appear, indicating that the SPAN 2 has been
accepted. Press
Enter
again to complete the calibration procedure.