Guidance and manufacturer’s declaration – electromagnetic immunity
Triangle is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that they are used in such an environment.
Immunity Test
IEC 60601
test level
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 8 kV contact
(1)
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
(1)
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 2 kV for power
supply lines
± 1 kV for input /
output lines
Mains power quality
should be that of a
professional healthcare
facility environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
±2 kV common
mode
± 1 kV differential
mode
±2 kV common mode
Mains power quality
should be that of a
professional healthcare
facility environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5% U
T
(2)
(> 95% dip in U
T
)
for 0,5 cycle.
40% U
T
(60% dip in U
T
) for
5 cycles.
70% U
T
(30% dip in U
T
) for
25 cycles.
<5% U
T
(> 95% dip in U
T
)
for 5 s.
< 5% U
T
(2)
(> 95% dip in U
T
) for
0,5 cycle.
40% U
T
(60% dip in U
T
) for 5
cycles.
70% U
T
(30% dip in U
T
) for
25 cycles.
<5% U
T
(> 95% dip in U
T
) for
5 s.
Mains power quality
should be that of a
professional healthcare
facility environment. If
the user of Triangle
requires continued
operation during a
power mains
interruption, it is
recommended that
Triangle be powered
with an uninterruptible
power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic field should be
at levels characteristic
of a professional
healthcare facility
environment.
Note:
(1)
A lock or reboot of the device with no permanent damage is acceptable
(2)
U
T
is the a.c. main voltage prior to application of the test level.
Summary of Contents for TRIANGLE
Page 2: ......
Page 3: ...EN AUDIOMETER TRIANGLE USER MANUAL ...
Page 8: ......
Page 14: ......
Page 24: ......
Page 30: ......
Page 34: ......
Page 47: ...IT AUDIOMETRO TRIANGLE MANUALE UTENTE ...
Page 52: ......
Page 58: ......
Page 68: ......
Page 74: ......
Page 78: ......
Page 91: ...DE AUDIOMETER TRIANGLE BEDIENUNGSANLEITUNG ...
Page 96: ......
Page 102: ...6 UDI Code ...
Page 108: ......
Page 114: ......
Page 120: ......
Page 124: ......
Page 138: ...42 ...
Page 139: ...FR AUDIOMETRE TRIANGLE MANUEL UTILISATEUR ...
Page 144: ......
Page 156: ......
Page 162: ......
Page 168: ......
Page 172: ......
Page 185: ...ES AUDIÓMETRO TRIANGLE MANUAL DEL USUARIO ...
Page 190: ......
Page 196: ...6 Código UDI ...
Page 202: ......
Page 208: ......
Page 214: ......
Page 218: ......
Page 231: ...PT AUDIÓMETRO TRIANGLE MANUAL DO UTILIZADOR ...
Page 236: ......
Page 242: ...6 Código UDI ...
Page 247: ...11 PT ...
Page 248: ......
Page 254: ......
Page 260: ......
Page 264: ......
Page 277: ...NL AUDIOMETER TRIANGLE GEBRUIKERSHANDLEIDING ...
Page 282: ......
Page 288: ...6 UDI code ...
Page 293: ...NL 11 Het instellingenscherm openen zie Hoofdstuk 5 ...
Page 294: ......
Page 300: ......
Page 306: ......
Page 310: ...28 ...
Page 323: ...PL PL AUDIOMETR TRIANGLE INSTRUKCJA OBSŁUGI ...
Page 328: ......
Page 340: ......
Page 346: ......
Page 352: ......
Page 356: ......