12
QA XXX
If the functions of the bed change, please send the bed immediately for a check-up according
to the maintenance chart, chapter 12.
Technical part
Invacare®
EC-Høng A/S is certified according to DS/EN ISO 9001 and ISO 13485 which ensu-
res that our customers are always supplied with
Invacare®
EC-Høng A/S products of uniform
quality.
Throughout the entire production process our materials and products are quality controlled
by the operators.
Furthermore there is a final test when the product is fully
assembled.
The operator carrying out the final test, which comprises
of checking all movable parts, motors and castors, issues a
QA number on the product thereby confirming it´s quality.
If the product does not correspond to the quality demands
of
Invacare®
EC-Høng A/S, it will be discarded.
If, contrary to our expectations, a problem should arise in connection with the delivered
product, please contact your
Invacare®
supplier.
3. Information
Please read the entire technical information section carefully before using or servicing the bed.
All indications of right and left are based on a patient lying on his back in the bed.
Please note that there may be sections in this user manual, which are not relevant for your
bed.
The
SB755
TM
is CE-marked in accordance with directive 93/42/EEC concerning Medical
Apparatus.
The
SB755
TM
is tested and approved by TÜV SÜD Product Service GmbH according
to EN 1970.
SB755
TM
electrical system is tested and approved in accordance with EN 60601-1 and
EN 60601-1-2 by UL Demko.
Max. patient weight: 200 kg.
The hand control, control unit and motors are protected according to IP 66. A lock cam must
be used on the control unit - if not,
Invacare®
EC-Høng cannot guarantee the IP protection.
Electromagnetic interference between the bed and other electrical products
can occur. To reduce or eliminate such electromagnetic interference,
increase the distance between the bed and the products or switch them off.
This medical bed can be used together with medical electrical equipment
connected intravascularly or intracardially, provided that following points are
respected:
The medical bed should be equipped with means for potential equalization
connection marked out by a symbol shown in the back of this manual.
Medical electrical equipment should not be fixed on the bed´s metallic
accessories such as side rails, lifting pole, drip rod, bed ends ect.
In addition, the medical electrical equipment power supply cord should be
kept clear of the accessories or any other moving part of the bed.
UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06.indd 12
2011-11-08 15:23:49
Summary of Contents for ScanBed 755
Page 17: ...20 UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06 indd 20 2011 11 08 15 24 03...
Page 51: ...20 UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06 indd 20 2011 11 08 15 31 17...
Page 119: ...20 UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06 indd 20 2011 11 08 15 45 55...
Page 153: ...20 UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06 indd 20 2011 11 08 16 12 29...
Page 187: ...20 UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06 indd 20 2011 11 08 16 37 47...
Page 221: ...20 UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06 indd 20 2011 11 08 16 45 21...
Page 255: ...20 UM_1518190_GB_DK_SE_NO_FI_DE_NL_FR_IT_PT_ver06 indd 20 2011 11 09 08 13 42...