SECTION 2—SAFETY INSPECTION/TROUBLESHOOTING
16
Wheelchair will not
drive.
Low tire pressure.
Motor engagement handles are dis-
engaged.
Batteries require charging.
Charger plugged in.
Circuit breaker tripped.
Inflate tires to the proper tire pressure.
Engage motor engagement handles.
Charge batteries. Make sure the setting on the
charger is correct.
Unplug charger from wall outlet before operating
the wheelchair.
Press the reset button. Refer to Resetting the Cir-
cuit Breaker on page 44.
Motors stall and
start up again.
Current rollback.
Stop driving and let electronics cool.
Motor “chatters”
or runs irregularly.
Electrical malfunction.
Contact Invacare.
Only one (1) drive
wheel turns.
Electrical malfunction.
One (1) motor is disengaged.
Contact Invacare for service.
Engage motor engagement handle.
Joystick erratic or
does not respond
as desired.
Electrical malfunction.
Controller programmed improperly.
Contact Invacare for service.
Reprogram controller (Refer to electronics man-
ual supplied with wheelchair).
Wheelchair does
not respond to
commands.
Poor battery terminal connection.
Have terminals cleaned.
Power indicator off
- even after
recharging.
Electrical malfunction.
Contact Invacare for service.
Wheelchair will not
drive.
Low tire pressure.
Motor engagement handles are dis-
engaged.
Batteries require charging.
Charger plugged in.
Circuit breaker tripped.
Inflate tires to the proper tire pressure (30 psi).
Engage motor engagement handles.
Charge batteries. Make sure the setting on the
charger is correct.
Unplug charger from wall outlet before operating
the wheelchair.
Press the reset button. Refer to Resetting the Cir-
cuit Breaker on page 44.
SYMPTOM
PROBABLE CAUSE
SOLUTIONS