120-00012
GasD 8000 Series User Manual
12/14/2020
6
To connect the probe to the instrument INLET first push down on the metal locking tab of
the inlet fitting then plug the tube adapter into the “IN” connector until you hear the locking
mechanism click into place (below, left). To remove the tube adapter, press down on the
metal locking tab on the top of the connector while pulling out on the tubing (below, right).
2.4 ANALOG OUTPUT CONNECTIONS
The GasD
8000 Series analyzer is equipped with 0-100 mV** and a 4-20 mA analog
output signals available at the 10 pin rear panel terminal block as shown in the table below.
For mV output, connect between terminals 1 or 2 (GND) as the negative reference and
terminal 7 (AOUT2) as the positive output as shown in blue in the table below. For 4-20
mA output, use terminal 3 (VCC) as the output and terminal 8 (AOUT1) as the return,
shown in green in the table below.
PROBE
TIP
PROBE
HANDLE
TUBING
TUBE
ADAPTER