background image

 

D-0134090-A – 2022/11  
VisualEyes™ - Instructions for Use - EN 

 

Page 

71

 

 

Top mount goggle

:

 

If using the top mount goggle, the images can be centered from the software by using the center eyes button 

from the eye tools menu, as described in section 3.6.3. 

 

Front mount goggle

:

 

If using the front mount goggle, use the center eyes button in the eye tools menu and manually adjust the 

camera in the goggle viewport. Ensure that the “UP” sticker on the camera is pointing upwards. 

 

 

EyeSeeCam goggle

:

 

If using the EyeSeeCam goggle, align the camera by rotating it in its ball-and-socket joint such that the eye is 

centered in the image. The camera can be rotated in three directions: yaw, pitch, and roll. Only rotate the 

camera in one of these three possible directions at a time. After the eye is centered in one direction, e.g. in 

the horizontal direction, choose the next rotation direction, e.g. the vertical direction. 

A possible alignment sequence to center the is shown below: 

 

 

 

  

 

Figure 3.6-1: Centering the pupil in the image for the EyeSeeCam goggle, with a sequence of three camera rotations 

 

 

3.6.2  Focus adjustment 

To optimize the pupil detection, the camera should be focused. This is done by turning the knobs/lens as 

described for the selected goggles in section 2.6.1, and shown in 

Figure 3.6-2

. Turn the knob clockwise or 

counterclockwise while watching the image on the screen. Stop turning the knob when the pupil outline and 

iris pattern is clear, even if the eye surroundings may be blurry at this point. The infrared reflections will be 

the smallest when the cameras are in optimal focus. With binocular goggles, each camera must be adjusted 

separately.  

 

NOTICE: 

If the image is not optimally focused, the eye tracker may fail to adequately record the eye 

movements, which in turn may lead to poor results.

  

 

 

Figure 3.6-2: Focus adjustment on the different VNG and vHIT goggles

 

 

 

3.6.3  Eye tools menu 

When touching or hovering the mouse over the eye images in the test screen, the 

Eye tools menu

 will 

appear. The Eye tools menu contains five buttons: 

 

1. 

Toggle Right Eye: 

Enable/disable image and eye tracking for the right eye. 

 

2. 

Click to center eyes: 

Centers the eyes in the images (only available with the top mount and front 

mount goggles). 

 

3. 

Click to select eye tracker: 

Allows the operator to select the eye tracker to be used. 

 

 

 

Summary of Contents for VisualEyes 505

Page 1: ...rotationa Instructions for Use EN VisualEyes Micromedical by Interacoustics Science made smarter D 0134090 A 2022 11...

Page 2: ...out notice No part of this document may be reproduced or transmitted in any form or by any means without a prior written permission of Interacoustics FireWire is a registered trademark of Apple Inc re...

Page 3: ...3 USB connection 18 Software installation 18 2 5 1 Demo patients 20 2 5 2 Uninstalling VisualEyes 21 Hardware components and installation 21 2 6 1 VNG and vHIT goggles 22 2 6 1 1 Side mount goggles 2...

Page 4: ...adjustment 70 3 6 1 Eye image centering 70 3 6 2 Focus adjustment 71 3 6 3 Eye tools menu 71 Calibration 72 3 7 1 Standard calibration 73 3 7 2 Torsion calibration 75 3 7 3 EyeSeeCam calibration 76 3...

Page 5: ...isualEyes system 99 Warranty and service 99 Rotational chair error messages 100 Cameras disconnecting 102 General technical specifications 103 Minimum computer requirements 103 Standards 103 Goggles 1...

Page 6: ...nuals on the manufacturer s official website Intended use The VisualEyes system is to be used by trained personnel only such as audiologists ENT surgeons physicians hearing healthcare professionals or...

Page 7: ...t rotary chairs The VisualEyes system is available in a number of different configurations with and without rotary chairs The standard configurations without rotary chairs are VisualEyes 505 VisualEye...

Page 8: ...Dynamic Visual Acuity o Gaze Stabilization Test VORTEQ Diagnostic VisualEyes 525 VORTEQ Diagnostic is an add on module including the VORTEQ 2nd Generation IMU for head position and velocity tracking...

Page 9: ...X Smooth Pursuit X Saccade X Ocular Counter Roll w torsional tracking X Saccadometry X EyeSeeCam vHIT X ADD ON VORTEQ Assessment Dynamic Visual Acuity Dix Hallpike Advanced Lateral Head Roll Gaze Sta...

Page 10: ...Orion Reclining Orion Reclining is a rotary chair that can be reclined for caloric and positioning testing Orion Comprehensive Orion Comprehensive is a rotary chair inside a chair enclosure containing...

Page 11: ...X X X X Sinusoidal Harmonic Acceleration SHA X X X X X X VOR Suppression X X X X X X Visual VOR X X X X X X Static SVV X X X X Dynamic SVV X X ADD ON VORTEQ Assessment Dynamic Visual Acuity Dix Hallp...

Page 12: ...ed major parts EyeSeeCam vHIT goggle and camera VisualEyes software Full HD room camera Cleaning cloth Introduction guides Certificates and reports e g test result report license certificate certifica...

Page 13: ...esktop PC configuration Included major parts Binocular USB goggles 7 port USB 3 0 hub w external power supply Included major parts Adult USBM2 1A Pediatric USBM2 1P mask for monocular USB camera 7 por...

Page 14: ...Goggle mounts for side mount and top mount goggles VORTEQ 2nd gen sensor Bluetooth dongle USB cable for charging wired connection Goggle mounts for side mount and top mount goggles Headband for Dynam...

Page 15: ...device to sale by or on the order of a physician 2 Do not disassemble or modify the product as this may impact on the safety and or performance of the device Always refer to qualified authorized pers...

Page 16: ...14 Do not service any part of the equipment while in use with a patient CAUTION used with the safety alert symbol indicates a hazardous situation which if not avoided could result in damage to the equ...

Page 17: ...y Such products will be marked with the crossed out wheeled bin The cooperation of the user is important to ensure a high level of reuse and recycling of electric and electronic waste Failing to recyc...

Page 18: ...operator can restart the test from the software Rotational chair has seat belt to secure the patient from falling during the tests Rotational chair tests do not respond to the RF remote as an additio...

Page 19: ...it Examine the cabinet and accessories for scratches and missing parts Report any concerns immediately Report any missing part or any mechanical or electrical malfunction immediately to your local di...

Page 20: ...tion EU 2017 745 Approval of the quality system is made by T V identification no 0123 Medical device Manufacturer Date of Manufacture Indicates a component is intended for one use or for use on a sing...

Page 21: ...16 This product contains a component Orion Comprehensive Auto Traverse chair of Class 2 Laser beam Hence do not look directly into the laser beam Keep dry Transport and storage temperature range Trans...

Page 22: ...VisualEyes This is especially important to consider in a server client setup with multiple installations of VisualEyes where it is recommended to upgrade all installations to the newest software PC se...

Page 23: ...and shut down the computer Make sure the power cable and the power supply for PC are disconnected Plug the PCI expresscard into a free PC slot and insert the power adapter F3 14 SATA 16 between the po...

Page 24: ...If a previous version of VisualEyes has been installed on the computer checking Migrate any existing VisualEyes system settings will transfer any existing system settings during the installation This...

Page 25: ...store it in a convenient place 2 5 1 Demo patients The VisualEyes software includes demo patients that you can import into OtoAccess Database for demos The demo patients are available at the followin...

Page 26: ...installer and control panel Hardware components and installation A VisualEyes system can depending on the configuration contain a lot of hardware components and it is important that these are all conn...

Page 27: ...justable Velcro head strap that comfortably secures the goggles to the patient The side mount goggles come with replaceable foam pads that are meant for single use as reuse of foam pads can cause cros...

Page 28: ...rizontal camera adjustment 3 camera focus adjustment right Hardware setup 1 Insert the camera s into the camera slot s on the goggle and tighten the fixation screws using the supplied screw driver unt...

Page 29: ...ect the other end of the USB cables to the supplied USB hub Ensure that the hub is powered through the external power supply If used with a rotary chair the goggles plug in to the connectors on the ch...

Page 30: ...igured as VisualEyes EyeSeeCam The camera is connected through USB and can be mounted into the spherical socket over the eye A laser stimulus projects from the goggles in the vHIT test Figure 2 6 7 Ey...

Page 31: ...ensing 6 Ensure that the calibration laser is in horizontal vertical alignment and adjust if necessary 2 6 2 External room camera All VisualEyes systems are delivered with an external room camera This...

Page 32: ...ase find the below example for better explanation Before calibration you can select the Calibration option which is shown with a white background After calibration you can select the START option whic...

Page 33: ...d installation instructions please refer to separate documentation for AquaStim or AirFx Setup in software 1 Connect the USB type B connector to the connector on the rear panel of the irrigator Connec...

Page 34: ...be model DLB7 2 or newer to be compatible with the system Figure 2 6 15 Digital Light Bar Hardware setup 1 Attach the Digital Light Bar to the tripod 2 Connect the USB cable to a USB port on the compu...

Page 35: ...a side mount goggle left and mounted on the headband for Dynamic Visual Acuity right The VORTEQ 2nd Generation IMU has four LEDs that describe the status of the device ON Flashes yellow when powered o...

Page 36: ...all the device software from Kvaser AB Check the box to always trust software from Kvaser AB then choose Install 2 6 6 1 Optotype Tests If VORTEQ Assessment is added to the software the stimulus must...

Page 37: ...tions for the hardware installation please refer to the separate installation instructions Hardware setup 1 Follow the instructions in the separate installation instructions for the specific rotary ch...

Page 38: ...position left and in reclined position right Reclining chair configuration The VNG goggles can be connected directly to the reclining rotational chair FireWire cards are used with FireWire side mounte...

Page 39: ...D 0134090 A 2022 11 VisualEyes Instructions for Use EN Page 34 Figure 2 6 21 Configuration of Nydiag 200 Reclining Chair Figure 2 6 22 Configuration of System 2000 Reclining Chair...

Page 40: ...p test by combining the VisualEyes system with Orion Auto Traverse Comprehensive or System 2000 Auto Traverse Comprehensive rotary chairs The chair is enclosed in a light tight booth The laser project...

Page 41: ...ear the operator This allows the operator to monitor the patient inside the booth via the observation monitor screen The video from the observation camera is not recorded or saved in the software Figu...

Page 42: ...D 0134090 A 2022 11 VisualEyes Instructions for Use EN Page 37 Figure 2 6 27 Configuration of Orion Auto Traverse Comprehensive Figure 2 6 28 Configuration of System 2000 Auto Traverse Comprehensive...

Page 43: ...crew the head support on the chair to remove it 2 Slide the hook through the seat belt slot of the safety seat so that a hook is available on each side of the infant seat 3 Secure the infant seat and...

Page 44: ...atible with the below tests if licensed Spontaneous Nystagmus Gaze Smooth Pursuit Saccade Optokinetic Step Velocity Sinusoidal Harmonic Acceleration Hardware setup 1 Ensure the EOG board is registered...

Page 45: ...slowly Watch the sine wave There should be two colored sine waves that are superimposed into one curve If the curves match there is no need to calibrate Figure 2 6 32 Sine Wave Validation If the curve...

Page 46: ...centered laterally on the chair base If the scale shows the chair is off center enter the current chair off center position under Current Lateral Position cm then select the direction the chair is fro...

Page 47: ...e Y Axis and observe if the laser moves in a smooth pattern Click Stop to stop the Sine Wave Test Under SVV Angle Test click Go to check that the SVV line is rendered Use the left right arrow buttons...

Page 48: ...drum on the Auto Traverser Comprehensive chair Safety checks Before beginning the test in the chair system certain safety tests must be done as described below Safety checks for for Orion Auto Traver...

Page 49: ...le to the EOG connector on the rear panel of the DataLink 2 Connect the USB type B connector to the USB PC port on the back of the DataLink and connect the other end to the computer or the USB hub The...

Page 50: ...000 Comprehensive Reclining chair To solve this change the board number by right clicking on the board in Instacal and click Change Board Configure instacal so the DataLink miniLAB 1008 is Board 1 and...

Page 51: ...be installed choose the default option to install the Universal Library Examples During installation Windows Security may request permission to install the device software from Measurement Computing...

Page 52: ...the default selections for installing the drivers 32 64 bit x86 Start Menu Shortcuts and GUI tools When asked for the destination folder for installing the drivers choose the default location of C Pr...

Page 53: ...hardware will then function using the same license as entered for the VNG goggle After registering and licensing the VNG goggle any additional hardware can be registered This can be done by restarting...

Page 54: ...D 0134090 A 2022 11 VisualEyes Instructions for Use EN Page 49...

Page 55: ...e used Wireless HDMI connections are not supported When setting up the TV this should be set to Computer or Game mode to correct for HDMI scaling and digital image processing potentially delaying the...

Page 56: ...during installation it is highly recommended to remeasure the display size Different versions of VisualEyes have used different measurement methods It is therefore important to validate the dimension...

Page 57: ...ing the sliders determining the maximum angle that can be used for the visual stimuili 2 9 3 2 Hardware delay In order to compensate for any potential delay for the presentation of the visual stimuli...

Page 58: ...elay can be found under System Default Settings Stimuli Clicking Reset Delay will set the hardware delay to 0 ms To adjust the hardware delay for the connected monitor for visual stimuli the below ins...

Page 59: ...e test click on the subtest 3 Go to Temporary Setup 4 From Temporary Settings click on Hardware Delay in the left panel 5 The suggested adjustment for the hardware delay will be presented as Current D...

Page 60: ...ured under System Default Settings DVA Stimulus The DVA Stimulus settings screen provides the separate screen measurements for the stimulus used for the DVA tests If the TV screen is too large for DVA...

Page 61: ...te language in the drop down selection menu see Figure 2 6 2 The software will have to be restarted when the program language is changed The languages available in the software are below English Unite...

Page 62: ...ase 2 11 2 Video recordings VisualEyes can record videos from the VNG or vHIT goggles as well as the video from the external room camera These video files will be linked to the specific patient and se...

Page 63: ...archiving PDF prints outside of OtoAccess Database This is possible from System Default Settings Print under PDF Configuration From here set Save PDF Document to ON and specify the storage location fo...

Page 64: ...onfiguration section in System Default Settings Print After following above instructions a new Create PDF button will appear in the Session Review after a completed session Clicking Create PDF will ge...

Page 65: ...ockwise to disengage the button DataLink optional Ensure that the DataLink is connected to power and connected to the computer through the USB connection Power on the device using the power switch on...

Page 66: ...rmation on testing see sections 3 8 and 3 9 4 Patient Sessions This will take the operator to an overview of previous patient sessions that can be reviewed in detail exported or printed For further in...

Page 67: ...D 0134090 A 2022 11 VisualEyes Instructions for Use EN Page 62 Figure 3 2 1 VisualEyes main screen...

Page 68: ...3 3 1 This menu is used for setting up the system components and accessories and general software settings such as language and Video Storage Location Figure 3 3 1 System Default Settings menu The Sy...

Page 69: ...or to the menu shown in Figure 3 4 1 This menu is used for managing the protocols in the system Protocols are composed of a list of tests and designated to be performed in a specific order at the pref...

Page 70: ...re the test can be divided in different sections for Sitting and Supine positions Table 3 4 1 Default protocols with VisualEyes systems VisualEyes 505 VisualEyes 515 VisualEyes 525 VORTEQ Assessments...

Page 71: ...sequence can be customized by the operator by adding or removing tests using the buttons below the list of tests The order of the tests can also be altered by clicking and holding on a test in the li...

Page 72: ...back on the goggles Confirm with the patient that there is no light leakage If the patient still sees light adjust the goggles position and tightness of the strap as needed 8 Headband straps should b...

Page 73: ...ither of the recline handles found on the sides of the chair frame and this is up to the operator to decide considering the test requirements The footrest should not be used as a step to enter the cha...

Page 74: ...the booth enclosure Unscrew the head support and remove it Slide the hook through the seat belt slot of the safety seat so that a hook is available on each side of the infant seat Secure the infant s...

Page 75: ...ately Ensure that the goggle cover is taken off while you place the goggle on the patient s face Adjust the strap for a snug fit For vision denied testing place the cover on the goggles Confirm with t...

Page 76: ...camera rotations 3 6 2 Focus adjustment To optimize the pupil detection the camera should be focused This is done by turning the knobs lens as described for the selected goggles in section 2 6 1 and...

Page 77: ...r is required to perform the calibration When using the Orion Comprehensive or Auto Traverse chairs the built in laser will be used for the calibration For EyeSeeCam it is recommended to always use th...

Page 78: ...impaired and not able to see the default target size Start the calibration by pressing Start Calibration 4 in Figure 3 7 1 Refer to the instructions in sections 3 7 1 3 7 4 for performing the calibra...

Page 79: ...et positions After completion of the calibration the check marks will turn green if the calibration values are within acceptable ranges Figure 3 7 3 If the calibration values are not acceptable the ch...

Page 80: ...n tracker is to be utilized in the dark with dilated pupils it is very important to perform the torsion calibration in the dark with dilated pupils If possible ask the patient to look slightly up this...

Page 81: ...bration must be performed Whenever EyeSeeCam is selected as input source the Head Calibration button will appear Figure 3 7 7 Clicking this will open the Head Calibration screen Figure 3 7 7 Head Cali...

Page 82: ...the impedance is measured as 10 kOhm or less the electrode will be marked as Good indicated by green coloring If the impedance is between 11 kOhm and 15 kOhm the electrode will be marked as Medium in...

Page 83: ...erformed by following the procedure in section 3 7 1 If the electrode traces for horizontal and vertical traces show drift the traces can be manually centered using the Center Traces button If the pat...

Page 84: ...e EN Page 79 Figure 3 7 11 Standard calibration with ENG The test screen When the system and the patient have been prepared the tests can be performed To enter a test session select the desired protoc...

Page 85: ...ins buttons for operator controls The buttons in the panel vary for each test Some of the most general controls are the following t Test menu opens the session tree allowing the operator to view the c...

Page 86: ...the test can be performed on an examination table reclining chair With the patient seated in the upright reclining chair remove the Orion or System 2000 head support and set it aside Unlock the chair...

Page 87: ...nd physically lift the chair back with the patient into an upright position Figure 3 9 2 Supine positional test 3 9 1 5 Caloric test Caloric testing is performed with the goggle cover on for vision de...

Page 88: ...y degrees above horizontal using the angle marker below the seat before performing the test Orion Reclining chair Recline the chair to thirty degrees above horizontal using the angle marker below the...

Page 89: ...The Velocity Step Test is vision denied test This involves rotating the patient in one direction at a constant velocity for several seconds then stopping the chair while recording continues The same p...

Page 90: ...izontal testing or up and down for vertical testing 3 9 2 2 Lateral Head Roll The Lateral Head Roll test is performed to identify horizontal canal BPPV and includes the 3D head model guide to assist i...

Page 91: ...otype character without shaking the head In the GST test the optotype size remains the same but the head velocity changes The head velocity is measured in degrees per second The results are plotted as...

Page 92: ...nt s head Practice a few impulses before recording This will also familiarize the patient with the stimulus Do not allow the hands to touch or move the goggles during head impulses as that movement wi...

Page 93: ...the operator performs the head impulse the software will display the head and eye movements in the appropriate impulse graph depending on the impulse direction If the head movement fits the swoosh ve...

Page 94: ...9 8 The analysis displayed is defined by the page navigation at the top of the screen Once the Lateral LARP and RALP tests have been performed the EyeSeeSix report can be created from the vHIT for Ey...

Page 95: ...ction This test resembles the lateral vHIT test and assesses the lateral semicircular canals A laser target is used as the suppression medium for the SHIMP test The vHIT goggle is placed on the patien...

Page 96: ...sitioned center laser dot The VOR gains should be similar in the vHIT and SHIMP tests However the pattern of saccades generated is different vHIT rarely generates compensatory saccades in normal patie...

Page 97: ...recordings are played synchronously from the Test Review screen The playback will begin by clicking on the play button in the playback menu As the test plays back a yellow circle will show the current...

Page 98: ...sualEyes system easily allows for reviews of current and or previous sessions for the selected patient After selecting the desired patient profile the operator can select the Patient Sessions button f...

Page 99: ...esults can be reviewed by selecting the test menu button in the top left corner of the test screen A drop down menu will appear with the Review Session button displayed at the bottom Figure 3 11 4 Rev...

Page 100: ...ower left corner of the Session Review screen to return to the main screen Review patient videos If video recording has been enabled during the patient sessions the videos may be reviewed after testin...

Page 101: ...atible video player Video of eye recordings no raw data Video of room recording By selecting the Enable Selection Mode in the left side panel it is possible to select videos to be deleted or exported...

Page 102: ...test protocol For a detailed description of each of these menus and the available settings refer to the Additional Information 3 13 2 Data export When you have completed your testing you can export r...

Page 103: ...ower switch on the back of the controller If the DataLink is used turn off the power using the power switch on the back of the device If the AirFx air irrigator or AquaStim water irrigator is used ver...

Page 104: ...een on PC can be cleaned by the microfiber lens cleaning cloth when the monitor is turned off If the touch screen needs further cleaning use a Sani Cloth AF3 disinfectant wipe to clean the surface All...

Page 105: ...lity and performance of the equipment if Assembly operations extensions readjustments modifications or repairs are carried out by authorized persons A 1 year service interval is maintained The electri...

Page 106: ...otary chair reclined position error applicable for Orion and System 2000 reclining chair If the Orion Reclining or System 2000 Reclining chair is reclined during a rotary chair test the software will...

Page 107: ...or go to the home screen and restarting the test will reset the watchdog timer This is indicated by the Emergency Stop lighting up again Cameras disconnecting If unusual behavior is experienced such a...

Page 108: ...indows 11 64 bit Standards IEC 60601 1 2005 AMD1 2012 Medical electrical equipment Part 1 General requirements for basic safety and essential performance IEC 60601 1 2 2014 Medical electrical equipmen...

Page 109: ...ge Environment Transport Temperature Storage Temperature Relative Humidity 20 50 C 0 50 C 30 80 Non condensing Interface FireWire USB 2 0 Cable Length 4 5m 3m for Orion Reclining Camera Configuration...

Page 110: ...Transport Temperature Storage Temperature Relative Humidity 20 50 C 0 50 C 30 80 Non condensing Interface USB 2 0 Cable Length 4 5m 3m for Orion Reclining 0 9m for Orion Auto Traverse Comprehensive C...

Page 111: ...15 35 C 30 90 Transport Storage Environment Transport Temperature Storage Temperature Relative Humidity 20 50 C 0 50 C 30 80 Non condensing Interface USB 2 0 Cable Length 4 5m 3m For Rotary Chair Ver...

Page 112: ...ive Humidity Ambient Pressure 15 35 C 30 90 98 104 kPa Transport Storage Environment Transport Temperature Storage Temperature Relative Humidity 10 50 C 10 35 C 30 80 Interface USB 2 0 Cable Length 2...

Page 113: ...used for EOG ENG measurements Is available as the DataLink through the EOG Accessory kit for VNG or built into the Orion Auto Traverse Orion Comprehensive chairs through the EOG Accessory kit for Orio...

Page 114: ...Use EN Page 109 DIGITAL LIGHTBAR Quick Description Digital Lightbar presenting visual stimuli for oculomotor testing Technical Specifications Interface USB 2 0 Dimensions L x W x H 83 8 x 8 9 x 4 4 c...

Page 115: ...on Environment Temperature Relative Humidity 5 40 C 30 90 Transport Storage Environment Transport Temperature Storage Temperature Relative Humidity 15 40 C 15 40 C 30 80 Rotation Control Mechanical In...

Page 116: ...rotary tests on top of standard VisualEyes Tests Step Rotation up to 200 s Sinusoidal Harmonic Acceleration 0 01 0 64 Hz VOR Suppression 0 01 0 64 Hz Interface USB USB Cable Length 3m Power Cable Leng...

Page 117: ...ing Rotation Control Software controlled Available rotary tests on top of standard VisualEyes Tests Step Rotation up to 350 s Sinusoidal Harmonic Acceleration 0 01 1 28 Hz VOR Suppression 0 01 1 28 Hz...

Page 118: ...ries PEDIATRIC OBSERVATION CAMERA Quick Description Camera mounted on a unipod that can be attached to the Orion Auto Traverse Comprehensive Can be used when doing measurements in babies who can t wea...

Page 119: ...fied by the manufacturer Otherwise degradation of the performance of this equipment could result in improper operation This equipment complies with IEC60601 1 2 2014 AMD1 2020 emission class B group 1...

Page 120: ...switch Interacoustics Emergency switch Goggle EyeSeeCam USB cable Interacoustics EyeSeeCam Anyone connecting additional equipment is responsible for making sure the system complies with the IEC 60601...

Page 121: ...environment in which radiated RF disturbances are controlled The customer or the user of the ORION can help prevent electromagnetic interferences by maintaining a minimum distance between portable and...

Page 122: ...e that of a typical commercial or residential environment Surge IEC 61000 4 5 1 kV Line to line 2 kV Line to ground 1 kV 2 kV Mains power quality should be that of a typical commercial or residential...

Page 123: ...culated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 Vrms IEC EN 61000 4 6 150kHz to 80 MHz 6 Vrms in ISM bands 150kHz to 80 M...

Page 124: ...eration If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the ORION b Over the frequency range 150 kHz to 80 MHz field strengths should be less...

Page 125: ...Returned according to agreement with Interacoustics Other Date Person Please provide e mail address or fax No to whom Interacoustics may confirm reception of the returned goods The above mentioned ite...

Reviews: