2.3.12 Initial network setup
The device is delivered with default network parameters (see
“3.4.3.6. Menu “Communication””
). The Ethernet/
LAN port is immediately ready for use after the initial commission. Default parameters are:
IP: 192.168.0.2
Subnet mask: 255.255.255.0
Gateway: 192.168.0.1
Port: 5025
DHCP: off
For wiring, i.e. the hardware connection to a network, seek support from your IT manager or any similar responsible
person. Network cable of common type (CAT5 or better) can be used.
In order to set up the network parameter to your needs, you have three options: the setup menu or the device’s
website or the software “EA Power Control“, which is available since April 2015 as a download or included with
your device. For the configuration in the setup menu please refer to
“3.4.3.6. Menu “Communication””
.
For the configuration through the device’s website or EA Power Control, you need the device to be connected to
a network or directly to a PC which can access the default IP 192.168.0.2.
►
How to do the network setup on the device website
1.
If the device display is in any kind of menu, leave menu to main display.
2.
Open the device website in a browser by entering the default IP (http://192.168.0.2) or the default host name
(http://Client, only possible if there is a running DNS in the network) into the URL box.
3.
After the website has been completely loaded, check the status field item “
Access
” to show the status “
free
”.
In case it shows different, the device is either already in remote control (
rem
) or blocked from remote control
(
local
). If it shows “
local
”, first remove the block. Refer to section
“3.5.2. Control locations”
to do that.
4.
If it says “
rem
” in the “
Access
” item, skip to step 4. Else enter command
syst:lock on
(attention! space
before
on
) into the
SCPI command
box and send with return key. Check if item “
Access
” in status field
changes to “
rem-eth
” (means: remote Ethernet).
5.
Switch to page
CONFIGURATION
(upper left corner) and set up the network parameters as well as the port
here resp. activate DHCP and submit the change with
SUBMIT
button.
6.
Wait a few seconds before testing the new IP by entering it in the browser’s URL box. Opening the website
again by using the host name is only possible after the device has restarted, because only then is the new
IP reported to the DNS.
2.3.13
Commission after a firmware update or a long period of non-use
After a firmware update, return of the equipment following repair or a location or configuration change, similar
measures should be taken to those of initial start up. Refer to
“2.3.11. Initial commission”.
Only after successful checking of the device as listed may it be operated as usual.