Innoquest SpotOn G2 Product Manual Download Page 8

 

13 

4.3.6  Once the meter has just finished 

a probing (before starting a new 
probing) the user can see 3-4 
types of data.  Pressing the 
START button will show the data 
in the following order.  If data 
logging is enabled then this data 
can also be viewed later through 
the use of Review Mode.   

 

SCREEN 1

 - The probing force 

average and temperature for the 
current reading called “NEW”.  The 
first 3cm and last 1cm of data are 
not used in computing this average.  
If more than 50% of the recorded force readings for the 
probing profile were at speeds greater or less than the middle 
three speedometer display segments (greater then +/- 20% of 
target speed) then the entire probing is rejected and the user 
is instructed to take a new one.  This keeps operator error 
from producing bad data sets.  If a Max Depth was set 
through the PC software for this product type then the 
average as described above will only be calculated on data 
points up to the set Max Depth. 
 

SCREEN 2

 - The multi-probe mean 

of both force and temperature for all 
probings since the meter was 
turned ON.  The standard deviation 
is also displayed for all readings 
included in the mean (shown as the Greek symbol sigma).  
The meter will only calculate the mean and std dev for up to 6 
probings.  If more than 6 probings are taken without turning 
the meter off/on then the mean will show as 0.  To reset this 
multi-probe averaging function turn the meter OFF then back 
ON. 

 

SCREEN 3

 - The min and max 

values are displayed as set through 
the PC software for the currently 
selected product type.  If the min 
and max are set to 0 and 9999 
respectively then this min/max screen will not show. 

 

This screen shows when the 
probing is completed. 

(whether by 

stopping the insertion and waiting or 
immediately withdrawing the probe)

 

14

SCREEN 3/4 & Up

 - Pressing the 

START button additional times will 
allow the user to cycle through the 
readings held in temporary memory 
for each 1 cm of probing depth.  
These readings are only recorded for computer download if 
the logging function is ON.  Once the max probing depth has 
been displayed the display will cycle back to the SCREEN 1 
in a continuous loop.  In this way you can always get back to 
a reading if you accidentally pressed the start button too 
many times. 

 

Pressing the MODE button at anytime will exit the review 
mode and ready the meter for another probing.  Once this is 
done the data described above is no longer available to view 
unless data logging was ON. 

 

 

 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for SpotOn G2

Page 1: ...Product Manual Innoquest Item 23920...

Page 2: ...of the user This meter is a low voltage battery powered device which should not cause electrical hazard to the user even when connected to a computer via a USB cable However please use caution not to...

Page 3: ...xtension Shows readings in either Grams or Pounds Integrated Temperature Sensor On board microprocessor calculates and displays the average force of a measurement the mean of a group of measurements t...

Page 4: ...omputer Interface USB port with special data interface cable included Battery Standard 9 volt alkaline battery 25 hrs continuous use expected battery life the meter will automatically shutoff after 2...

Page 5: ...ta file so the user can track where the probing came from with his her notes When the logging function is off the N number will only show as It is possible to turn the logging function on and off as m...

Page 6: ...on to exit setup mode Unless no PC is available or desired for operation it is best to set data logging ON 10 4 2 4 RAW MODE This mode allows the user to see the raw readings from the different sensor...

Page 7: ...the indicator must be centered ooOoo If the speed is too fast then the indicator moves to the right oooOo If the speed is too slow then the indicator moves to the left oOooo For reference Ooooo 60 of...

Page 8: ...in the mean shown as the Greek symbol sigma The meter will only calculate the mean and std dev for up to 6 probings If more than 6 probings are taken without turning the meter off on then the mean wi...

Page 9: ...windows compatibility mode and run as XP compatible USB Cable Installation The meter has a data port for connection to a computer via a special USB cable that was provided with the meter This cable r...

Page 10: ...the meter even after a download or after changing the meters battery The ONLY way to clear the data from memory is described in section 5 2 3 Data File Example Opened in a Spread Sheet Program 18 5 2...

Page 11: ...f no min max is desired to show on the meter Valid ranges are 25 to 9998 grams or 0 to 22 00 lbs Configuration This is a pull down menu that allows the user to select one of the custom named configura...

Page 12: ...matically show a centered position until the probe tip reaches a depth of 2 cm This allows time for the user to enter the surface of the product and then develop some consistency to their speed before...

Page 13: ...ir Service Policy Product returned to Innoquest for repair or service must follow the guidelines set forth as follows Return of the product for warranty or service repair will be the responsibility of...

Reviews: