4.2.3 SETUP
MODE
Once in Setup Mode press the START button to select
which parameter to change.
a. UNITS screen allows the user to change the
measurement units of the meter between Metric
and Imperial. Press the MODE button to enter
the units screen and then change the units with
the START button. Note there are two units
screens; one for force/depth and one for
temperature. When done press the MODE
button to exit Setup Mode. The measurement
units can also be changed with the PC software.
Either method has the same effect.
b. DATA LOGGING screen allows the user to select
if the meter is configured to save probing data to
its memory for later download to a computer or
review. Press the MODE button to enter the data
logging screen and then change the setting with
the START button. When done press the MODE
button to exit setup mode. Unless no PC is
available or desired for operation it is best to set
data logging ON.
10
4.2.4 RAW MODE
This mode allows the user to see the raw readings from
the different sensors on the meter. While in this mode
the user can press the START button to scroll through
several other screens used for diagnostic or
troubleshooting work. All of these screens are read only
so no parameters can be accidentally changed. Exit raw
mode at anytime by pressing the MODE button.
4.2.5 REVIEW MODE
This mode allows the user to see data from previous
probings (only available if data logging is turned on).
Pressing the START button again will allow the user to
scroll down through the previous probings. Once the
probing of interest is reached then press the MODE
button to select it. At this point the START button will
allow the user to see the data from this probing including
each 1 cm of recorded data by continuing to press the
START button. When done press the MODE button to
return to the review mode.