5
3.0 Specifications
Force:
0-22 lbs (0-9,999grams) working range, 0.01 lbs (4 g)
resolution, +/- 2% accuracy, Overload protection to 62 lbs (28
kg)
Depth:
0-28 in (0-70 cm) working range, 0.1 inch (0.1cm) resolution,
+/- 0.25” (0.6 cm) accuracy (Accuracy is much higher at
closer ranges and falls off with distance)
Temperature:
32.0-99.9*F (0.0-37.7*C) range, 0.1*F (0.1*C)
resolution, +/-0.5*F (0.3*C) accuracy, time constant approx
20 sec
Data Measurement:
Records and displays force reading for every 1
cm (0.4in) of depth up to 70cm (27.6in). (Depending on shaft
length used)
Memory Capacity:
Up to 99 probings
Multi Reading Average:
Up to 6 probings can be averaged on the
meters display (also calculates standard deviation)
Computer Interface:
USB port with special data interface cable
(included)
Battery:
Standard 9 volt alkaline battery, 25 hrs continuous use
expected battery life (the meter will automatically shutoff after
2.5 minutes of non-use to save battery life)
Environmental:
35-120*F (2-49*C) / 0-95% RH / Indoor operational
conditions (meter is dust tight but not waterproof)
6
4.0 General Operation
Note:
See Demo Video on www.YouTube.com
(Search SpotOn Consistometer)
4.1 Battery Installation
The meter requires a standard 9 volt alkaline battery. Use of
rechargeable batteries is not recommended since the meter won’t
read their remaining life correctly due to voltage differences. The
battery compartment is under the meter’s back cover. Remove the
back cover by first removing the (4) four Philips screws located in its
corners. Changing the battery does NOT erase any of the meter’s
memory; data or settings.