Chapter 7
Maintenance Menu
39
•
The splicer will show up two values on the screen after each arc calibration. If
the value shown on the right-hand side meets 11±2, the message “Operation
Complete” will be shown on screen. Otherwise, set prepared fibers on the splicer
again and redo arc calibration until the message “Operation Complete” is
displayed.
Electrode setting
Set the electrode change warnings.
INNO Instrument
recommends replacing the
electrodes every 3500 discharge to ensure the best splice results.
•
Select [Electrode Setting] in [Maintenance Menu].
•
Set the electrode caution and electrode warning.
Update software
•
Select [Update Software] in [System Setting].
•
Connect the USB driver with the copy of upgrade files to the USB port.
•
Press “OK”, and then the splicer will process the update automatically. Upon
completion, the splicer will restart automatically.
Parameter
Description
Electrode Caution
If actual arc discharge exceeds the set value, a message “Caution! Replace
Electrodes” is displayed when turning on the splicer.
INNO Instrument
recommends that this parameter should be set to “3500”.
Electrode Warning
If actual arc discharge exceeds the set value, a message “Warning! Replace
Electrodes” is displayed when turning on the splicer.
INNO Instrument
recommends that this parameter should be set to “5000”.