Chapter 3
Basic Operation
23
Note: The fibers are checked automatically when you press “Set” button. The
splicer automatically focuses the fibers and checks for damage or dust particles.
Splicing
•
Select any appropriate splice mode.
•
Start splicing by pressing [SET] button.
Note: If the splicer is set as “Auto mode”, splicing will start automatically once
the windproof cover is closed.
How to protect the splice
After splicing, put the fiber with heat-shrink sleeve into the heat oven. Press [HEAT]
button to execute heat-shrink process to strengthen the splice point.
Heating procedure
•
Open the heat oven lid.
•
Lift the left and right fiber holders on the splicer. Hold the heat-shrink sleeve
(previously placed onto the fiber). Lift the spliced fibers and hold them taut. Then
move the heat-shrink sleeve to the splice point.
•
Place the fiber with heat-shrink sleeve in the heat oven clamp.
•
Press [HEAT] button to start heating. Upon completion, the heating LED
indicator will go off with buzzing.
Dust on fiber
Chip
Tag
Large cleave
angle