Chapter 7
Maintenance Menu
37
•
Execute [Diagnostic test], then the following checks will be made.
Dust check
Through image acquisition, the splicer detects dust and contaminants on the
splicer, camera, and objective lenses that may result in improper splicing. This
function checks the optical path for the presence or absence of contaminants,
and judges whether they will affect the quality of fiber splicing.
Operation procedure
•
Select [Dust check] in [Maintenance menu].
•
If fibers are set in the splicer, remove them and press [SET] again to start the
dust check.
•
If dust is detected during the dust check process, the message “Find Dust On
The Lens” will be displayed on screen. Press the return button twice, and the
location of contamination will be shown on the screen. Then clean the objective
lenses, and [Dust check] until the message “Operation Complete” is shown on
screen.
Note: If contamination still exists after cleaning the objective lenses, please
contact your nearest sales agency.
Motor calibration
Motors are adjusted at the factory before shipping. However, settings may be
changed in various reasons. This function automatically calibrates the speed of 6
motors.
Parameter
Description
1
LED Calibration
Measure and adjust the brightness of LED.
2
Dust Check
Check the optical path for dust or dirt and judges whether they
disturb fiber observation. If contamination exists, press the return
button twice to display the location.
3
Motor Calibration
Automatically calibrate the speed of 6 motors.
4
Arc Calibration
Automatically calibrate the arc power factor and fiber splicing
position.