CLOVER A1c Self • 54 •
16. TROUBLESHOOTING
Hi
mg/dl
ou
Hi
mmol/L
Please double check the accuracy of the
test strip by performing control solution
test and re-test with a fresh blood sample.
The blood glucose
test result is higher
than 600mg/dl or
33.3mmol/L
Unplug the power adaptor and reconnect.
If the problem still persists,
contact your local representative.
There is a problem
with the power
adaptor or the
analyzer itself
The
Analyzer
does not
power on
>14%
Lo
mg/dl
ou
Lo
mmol/L
Please use the check cartridge to
validate the analyzer and re-test the
blood sample.
The HbA1c test
result is higher
than 14.0%
(130mmol/mol)
Please double check the accuracy of the
test strip by performing a control solution
test and re-test with blood sample.
The blood glucose
test result is lower
than 10mg/dl or
0.6mmol/L
ERROR CODE
CORRECTIVE ACTION
DESCRIPTION
CLOVER A1c Self • 55 •
17. SPECIFICATION
Storage Temperature
10 ~ 35 (50 ~ 95 )
Option
Glucose
Test Module
PC Connection Cable
Thermal Printer
Barcode Scanner
Relative Humidity Range
10% ~ 90%
Operating Temperature
17 ~ 32 (63 ~ 90 )
Dimensions/Weight
200*200*139(mm) / 1.4kg
Power Required
DC 9 V-1.5 A
Wavelength
415 nm (normal)
Memory Capacity
200 test results
Reading Time
5 minutes
Test Range
4.0~14.0%
Sample Volume
4
S a m p l e T y p e
Capillary whole blood,
Venous blood with anticoagulant