6 | Operation
INFICON
72 / 108
XL3000flex-Operating-instructions-jina83en1-07-(2105)
• Control unit
Control unit
Main Menu > Functions > Data >
Recorder > Delete > Delete files
6.7 Copying measurement data, deleting
measurement data
The measurement data can be saved to a USB stick, see Device setup [
14].
• "Main Menu > Functions > Data > Recorder > Copy > Copy files"
• "Main Menu > Functions > Data > Recorder > Delete > Delete files”
6.8 Updating the software
Software updates from INFICON are installed with the aid of a USB flash drive. The
update function of the device can be found under "Functions > Data > Update".
An update is possible,
• if one or several updates are available on the USB-Stick, but only one update per
type at most (control unit, MSB box, I/O module),
• if these parts are also connected free of disturbances and have an update
function.
The corresponding buttons in the update menu such as "Control Unit", "MSB Box",
and "I/O Module" are active and can be activated individually.
NOTICE
Aborted connection
Data loss due to an aborted connection
► Do not switch off the device and do not remove the USB flash drive while the
software is being updated!
► Switch the device off and back on after a software update has taken place.
6.8.1 Updating the software of the control unit
The software is included in two files named Handset_IFC_Vx.xx.xx.exe and
Handset_IFC_Vx.xx.xx.key.
1
Copy the file into the main directory of a USB flash drive.
2
Connect the USB flash drive to the USB port on the device.
3
Select: Functions > Data > Update.
Summary of Contents for XL3000flex
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Page 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 107: ......
Page 108: ......