INFICON
Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105)
55 / 108
6.3 Settings for the measurements
6.3.1 Select gas type (mass)
The machine, calibration and sniff factor are dependent on the configured mass and
are saved in the mass spectrometer module.
2
3
4
H
2
(Hydrogen, forming gas)
3
He or deuterated hydrogen (HD)
4
He (Helium) (factory setting)
Control unit
Main menu > Settings > Mass
LD protocol
Command 506 with value 2 (3, 4)
ASCII protocol
Command *CONFig:MASS 2 (3, 4)
6.3.2 Display equivalence leak rate for other gas
If you measure with the test gases helium or hydrogen, but want to display another
gas with its leak rate, use a correction factor for the test gas used.
1
2
Fig. 3: Measurement screen with displayed equivalence leak rate and configured favorites key
1 Display of gas name and equivalence factor
2 Favorites key for fast configuration of "gas equivalent selection" after set up,
see "Assigning favorite buttons [
44]"
Summary of Contents for XL3000flex
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Page 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 107: ......
Page 108: ......