Ambient conditions
XL3000flex, XL3000flexRC
Permissible ambient temperature (during
operation)
10 °C ... 40°C
Max. altitude above sea level
2000 m
Max. relative humidity ≤ 31°C
80 %
Max. relative humidity > 31°C
50 %
Storage temperature
-20 °C ... 60°C
Degree of contamination
2
4.8 Factory settings
The following table shows the factory settings in the "Sniffing" mode.
Parameter
Factory setting
AO upper limit exp.
1 x 10
-5
Equivalent factor mass 2 (H2)
1.0
Equivalent factor mass 3
1.0
Equivalent factor mass 4 (He)
1.0
Molar mass of the equivalence gas (test gas
mass 2 (H2))
2.0
Molar mass of the equivalence gas (test gas
mass 3)
3.0
Molar mass of the equivalence gas (test gas
mass 4 (He))
4.0
Operation mode
XL Sniffer Adapter
Bus module address
126
Pressure capillary blocked
(Low Flow)
0.2 mbar
Pressure capillary broken
(Low Flow)
0.6 mbar
Clogged pressure capillary monitoring
- with XL Sniffer Adapter (High Flow)
150 mbar
Broken pressure capillary monitoring
- with XL Sniffer Adapter (High Flow)
400 mbar
Pressure unit (interface)
mbar
Emission
On
Filter leak rate threshold
1 x 10
-10
Summary of Contents for XL3000flex
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Page 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 107: ......
Page 108: ......