Function
State:
Description
Measure
closed:
open:
Measuring (ZERO is possible, all trigger outputs switch depending on the leak
rate.)
Standby or emission disabled (ZERO is not possible, all trigger outputs will
return "Leak rate threshold value exceeded".)
Standby
closed:
open:
Standby (ZERO is not possible, all trigger outputs will return “Leak rate
threshold value exceeded”.)
Measuring (ZERO is possible, all trigger outputs switch depending on the leak
rate.)
SNIF
closed:
open:
SNIF
VAC
Error or
warning
closed:
open:
Error or warning
No error or warning
Gas ballast
closed:
open:
Gas ballast is active
Gas ballast is inactive
Calibration
leak open
closed:
open:
calibration leak is active
calibration leak is inactive
CAL stable
closed:
open:
Calibration completed with calibration leak (see "Time and general
60]")
Assignment not stable or calibration is inactive
Cathode 2
closed:
open:
Cathode 2 is active
Cathode 1 is active
6.2.24 Settings for bus module BM1000
Address of bus module
Setting the bus module address. (Node address with Profibus, MACID with
DeviceNet)
0 ... 255
Control unit
Settings > Set up > Interfaces > Bus module > Address
LD protocol
326
ASCII protocol
–
6.2.25 Operation mode “Sniffing”
The device has the sniffer mode with a high flow rate.
For the XL3000flex , only this operation mode makes sense.
Summary of Contents for XL3000flex
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Page 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Page 107: ......
Page 108: ......