INFICON
Operation | 6
XL3000flex-Operating-instructions-jina83en1-07-(2105)
35 / 108
6.2.2 Setting date and time
Setting the date
Format: DD.MM.YY
Control unit
Main Menu > Settings > Date/Time > Date
LD protocol
Command 450
ASCII protocol
*HOUR:DATE
Setting the time
Format: hh: mm
Control unit
Main Menu > Settings > Date/Time > Time
LD protocol
Command 450
ASCII protocol
*HOUR:TIME
6.2.3 Selecting a unit for the leak rate
Leak rate unit display
Selecting the leak rate unit in the display for sniffing
0
mbar l/s (factory setting)
1
Pa m
3
/s
2
atm cc/s
3
Torr l/s
4
ppm
5
g/a
6
oz/yr
7
sccm
8
sft
3
/yr
Control unit
Main Menu > Display > Units (Display) > Leak Rate Unit SNIF
LD protocol
Command 396 (sniffing)
ASCII protocol
Command *CONFig:UNIT:SNDisplay