56
7 Warning and error messages
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
HLD
6000
, k
ina4
3
e1-
a
, 140
8
E133
EEPROM error in sniffer line
The EEPROM in the sniffer line is defective
Contact INFICON customer service.
E134
Protocol error when communicating
with the sniffer line
The interface to the sniffer line does not function
correctly.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
E135
Checksum error when communicating
with the sniffer line
– The interface to the sniffer line does not function
reliably.
– Possible electrical sources of disturbance close
by.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– Eliminate sources of disturbance
– If the problem still exists, contact INFICON customer
service.
E136
No response from the sniffer line
The interface to the sniffer line does not function
correctly.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
E137
Resetting the sniffer line
– The sniffer line has reset itself.
– Possible electrical sources of disturbance close
by.
– The sniffer line is not connected correctly.
– Eliminate sources of disturbance
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
E138
Checksum error in EEPROM on the sniffer
line
The EEPROM in the sniffer line has false data or is de-
fective.
Contact INFICON customer service.
E139
Empty EEPROM in the sniffer line
The EEPROM in the sniffer line has no data or is de-
fective.
Contact INFICON customer service.
W140 Acceleration of the probe was perma-
nently too high within the last 5 minutes
– The sniffer line is not lying still.
– Defect in the sniffer line.
– Place the sniffer line on a steadier surface.
– Contact INFICON customer service.
W151 No connection to operating unit
Internal connection problems between the basic
unit and the operating unit.
Contact INFICON customer service.
W163 COOL-Check is not connected
COOL-Check is not connected or not connected cor-
rectly.
– Connect a COOL-Check, otherwise confirm the warn-
ing and calibrate externally.
– Check the COOL-Check connection.
W164 Checksum error in COOL-Check
– The COOL-Check is not connected correctly.
– The COOL-Check or the basic unit is defective.
– Check the COOL-Check connection.
– Check the connection of the COOL-Check to basic unit
(separate and reconnect; try another COOL-Check, if
possible).
– If the problem still exists, contact INFICON customer
service.
Table 10: Warning and error messages (Contin.)
No. Message
Possible
error
sources
Troubleshooting
Summary of Contents for HLD6000
Page 22: ...22 4 Description Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 28: ...28 5 Installation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 54: ...54 6 Operation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 71: ...10 Appendix 71 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 72: ...72 10 Appendix Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 74: ...74 Index Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 75: ...75 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...