Changing the filter block
►
Change the filter block with the integrated fine filters at least 1 x per month.
The filter block can be found at the base of the sniffer tip,
see "Fig. 16: The filters in the sniff-
. To replace the filter block, carry out the following:
1
Switch the device off.
2
Loosen the cap nut at the lower end of the sniffer tip and separate the sniffer tip from
the probe handle.
3
Pull the filter block out of the base of the sniffer tip.
4
Insert a new filter block into position.
5
Attach the sniffer tip back onto the probe handle and tighten the cap nut at the lower
end of the sniffer tip.
Replacement filter block
see "10.1 Accessories and spare parts", page 67
8.3
Sending for maintenance or repairs
You can send your device to INFICON to have it maintained or repaired. For further infor-
mation regarding this topic
see "9.2 Sending in the device", page 65
.
Summary of Contents for HLD6000
Page 22: ...22 4 Description Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 28: ...28 5 Installation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 54: ...54 6 Operation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 71: ...10 Appendix 71 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 72: ...72 10 Appendix Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 74: ...74 Index Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 75: ...75 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...