7 Warning and error messages
57
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
HLD
6000
, k
ina4
3
e1-
a
, 140
8
E165
Timeout of EEPROM COOL-Check
– The COOL-Check is not connected correctly.
– The COOL-Check or the basic unit is defective.
– Check the COOL-Check connection
– Check the connection of the COOL-Check to basic unit
(separate and reconnect; try another COOL-Check, if
possible).
– If the problem still exists, contact INFICON customer
service.
W166 Audio amplifier is faulty
Error in the internal audio amplifier
Contact INFICON customer service.
2xx voltage error
W220 V24V outside the range
– Line error on the M12 socket or the module con-
nected there.
– Internal defect.
– Check the connections.
– Contact INFICON customer service.
W230 V3.3V outside the range
Internal defect.
Contact INFICON customer service.
W240 V12V outside the range
– Defect on the sniffer line
– Internal defect.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W241 Voltage -12V outside the range
– Defect on the sniffer line.
– Internal defect.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W250 V5V outside the range
Internal defect
Contact INFICON customer service.
W253 Incorrect sniffer line voltage
Defect on the sniffer line
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
3xx Measurement system error
W324 Lamp current outside the range
– Sniffer line connection or infra-red source in the
sniffer line is defective.
– Internal defect in the basic unit.
– Check the connection of sniffer line to basic unit
(separate and reconnect; try another sniffer line, if
possible).
– If the problem still exists, contact INFICON customer
service.
W363 Sensitivity too low
– Optical cell is contaminated with water vapor.
– The optical cell is dirty
– The sensor in the sniffer line is defective.
– Depending on the quantity of water inside the opti-
cal cell, let the HLD6000 run between 1 minute and
two hours to clean the optical cell.
– If the problem still exists, contact INFICON customer
service.
W364 Sensitivity too high
Defect in the sniffer line
Contact INFICON customer service.
Table 10: Warning and error messages (Contin.)
No. Message
Possible
error
sources
Troubleshooting
Summary of Contents for HLD6000
Page 22: ...22 4 Description Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 28: ...28 5 Installation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 54: ...54 6 Operation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 71: ...10 Appendix 71 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 72: ...72 10 Appendix Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 74: ...74 Index Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 75: ...75 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...