6 Operation
33
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
HLD
6000
, k
ina4
3
e1-
a
, 140
8
6.2.3
Display on the sniffer line
An LED is mounted on the sniffer handle where you can read the different operating
modes,
Fig. 12: Sniffer line display
a
see "6.4.6.1 Time and type of calibration", page 43
b
Status LED
c
Filter holder on the sniffer tip
d
Designation of the gas that the sniffer line sniffs.
Table 7: Signals of the LED on the handle grip
Operating status
LED
Not connected
Off
No communication
Blue, flashing
Run up
Blue
Standby
Blue, flashing
Measuring mode
Green
Leakage rate >40 % of the setpoint
Yellow
Leakage rate >100 % of the setpoint
Yellow, flashing
Calibration
Blue, flashing
Error/Warning
Red, flashing
Error/Warning and leakage rate > 40 % of the setpoint
Red/green, flashing
Error/Warning and leakage rate > 100 % of the setpoint
Red/yellow, flashing
a
b
c
d
Summary of Contents for HLD6000
Page 22: ...22 4 Description Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 28: ...28 5 Installation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 54: ...54 6 Operation Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 71: ...10 Appendix 71 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 72: ...72 10 Appendix Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 74: ...74 Index Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...
Page 75: ...75 Translation of the original operating instructions HLD6000 kina43e1 a 1408 ...