tinb60e1 (2020-07)
11
Relative humidity
0 … 95%, non-condensing
Use
Operating
altitude up to 2500 m (8200 ft.)
Storage
altitude up to 12500 m
(41000 ft.)
Mounting orientation
horizontal recommended
(orientation has no effect on
measurements below 1.3 mbar
(1 Torr))
DeviceNet Interface
Device type
vacuum gauge/pressure gauge
device
Adjustable parameters
setpoints, engineering units of
measure, vacuum and atmos-
phere calibration
Messaging
polled I/O and explicit
Baud rates
125K, 250K, or 500K, adjust-
able via rotary switch
Supply voltage
12 … 26 V (dc), 0.22 A, 2.4 W
protected against power rever-
sal and transient over-voltages
Setpoint relays
two, single-pole double-throw
relays (SPDT), 1 A at 30 V (dc)
resistive, or V (ac) non-induc-
tive
Electrical connection
9-pin D-sub male used for
setpoint relays
5-pin Micro for power and
DeviceNet interface