tinb60e1 (2020-07)
35
RLY:
[Factory default = DISABLED]
The factory default setting for the relay 1 is disabled. Select this
menu to change the state of relay 1 to ENABLED.
LO TRIP
[Factory default = 7.00 E+02 TORR]
This setpoint corresponds to the turn on point for Relay 1. Relay
1 will turn on when the pressure drops below this setting. If you
are unable to increase the value of LO TRIP (PGE500 respond-
ing with message "LO MORE THAN HI"), you must first access
the HI TRIP menu below and increase that value to a number
higher than the value of the LO TRIP you are trying to set. Note
that at the bottom of the screen the HI TRIP value is also dis-
played (but cannot be changed) while in this menu selection.
HI TRIP
[Factory default = 7.50 E+02 TORR]
This setpoint corresponds to the turn off point for Relay 1. Relay
1 will turn off when the pressure rises above this setting. If you
are unable to decrease the value of HI TRIP (PGE500 respond-
ing with message "LO MORE THAN HI"), you must first access
the LO TRIP menu above and decrease that value to a number
lower than the value of the HI TRIP you are trying to set. Note
that at the bottom of the screen the LO TRIP value is also dis-
played (but cannot be changed) while in this menu selection.
RELAY 2
Select RELAY 2 to configure setpoints for relay 2 and the follow-
ing submenus will appear:
RLY:
[Factory default = DISABLED]
Same instructions as relay 1 RLY above.
LO TRIP
[Factory default = 7.00 E+02 TORR]
Same instructions as relay 1 LO TRIP above.