tinb60e1 (2020-07)
51
7.7
Configuring Polled I/O data format
The PGE500 can provide input polled pressure data in UINT in-
teger or REAL floating point formats, with or without BYTE status
data. If a polled connection is established before this message,
remove and reestablish the connection. The status byte indi-
cates Warning in bit 5 (reading lower than zero vacuum, calibra-
tion needed) and Alarm in bit 1 (sensor failure). Default configu-
ration is 1 byte status and 4 bytes floating point pressure.
Format
Set
Service
Assembly
Class
In-
stance
Attri-
bute
Data
2 bytes integer
pressure
10 hex
4
0
65 hex
1
1 byte status and
2 bytes integer
pressure
10 hex
4
0
65 hex
2
4 bytes floating
point pressure
10 hex
4
0
65 hex
4
1 byte status and
4 bytes floating
point pressure
10 hex
4
0
65 hex
5