IDS 410/422 User's Manual
24
Parameter Functions
Fn.
No
Function Name
Password Level
59
Display Calibration Audit Number
0
60
Calibrate Scale - deadload first
3
61
Calibrate Scale - span first
3
62
Calibrate Deadload Only
3
63
Adjust Gain Calibration
3
64
Configure Passwords
2
65
Configure Print Formats
2
66
Set Time and Date
2
67
Display Operation Parameters
0
68
Print Operation Parameters
0
69
Diagnostic Tests
0
70
Initialize Operation Parameters to factory defaults.
3
71
Configure Event Monitor
2
72
Function Editor
2
73
Select Mode of operation (0-10)
2
74
Configure timers
2
76
Single-step Scale Basic functions
2
78
Erase ID Memory
2
79
Enable Multi-point Linearization
2
80
Multi-point calibration function
2
Scale Parameters
The scale parameters configure IDS 410/422 for the weigh platform or load cell that it is
connected to. The preceding table titled "SCALE PARAMETERS” lists the scale
parameters and their factory settings. Use the "Field Setting" column to record any changes
made to the factory default values.
Many of the following parameters are entered in terms of "graduations". The full-scale
graduations can be found by dividing the full-scale weight by the count-by weight. For
example, if the scale is configured for 500.00kg x 0.05kg, the full-scale graduations is
500.00/0.05 = 10,000 graduations to full scale.
The numbers in parenthesis are the default factory settings.