IDS 410/422 User's Manual
10
Setup Guide
Setup the IDS 410/422 in the following order:
Initialize to Factory Defaults
The IDS 410/422 is initialized at the factory. It must be re-initialized if the program memory
(EPROM, U4 , RAM U5 ), the parameter memory (EAROM, U12) is replaced or if the optional
clock chip in being installed for the first time. The Initialize function clears the memory and sets
the configuration parameters to their default values. Use parameter function 70 initallize system.
When the display prompts “rESEt0” enter
10809
code as described in chapter 6 to initialize the
IDS 410/422.
NOTE: Configuration and calibration and must be performed after the IDS 410/422 is
initialized
Setup & "Quick" Calibration Guide
Set the configuration parameters BEFORE calibration. Chapter 5:
SET OPERATION
PARAMETERS
details the parameters that are available for configuration. To quickly get the
IDS 410/422 up and running, the following Quick Calibration Guide lists only the mandatory
parameters that must be defined.
The parameters are accessed by holding the GROSS/NET key down and then pressing the
UNITS key. The numeric display prompts "CFG xx", where xx is the currently selected
parameter. The alternate functions (
clear, item, +, -, enter
) of the keyboard are now active.
Press the
item
key to select a digit, press the (
+
) key to advance the CFG number by 1, and press
the (
-
) key to decrement the CFG number by 1. Press the
enter
key to begin configuration of the
displayed parameter number.
Parameter
Purpose (Factory Setting)
CFG 1
Load Cell mv/V (3)
CFG 11
Decimal Point (0 [none])
CFG 12
Count-by (1)
CFG 17
Full Scale Graduations (10,000)
CFG 18
Overload Graduations (10,200)
CFG 73
Select Mode of Operation
NOTE: Read the first section of chapter 5:
Access Operation Parameters
for detailed
instructions on how to access and set the parameters.