- Check pneumatic system (visual inspection)
Orientation
To ensure trouble-free operation, periodic checks (visual inspection) of the
pneumatic system are necessary.
- Check oil level at oiler (optional).
- Check system pressure and sealing air settings.
- Check supply and fluid lines.
- Check silencers.
- Drain condensate (not applicable to auto-drain).
- Check pressure setting on the pressure differential monitoring pressure gauge.
Example: Pneumatic maintenance unit G200.2 by Festo
Due to different requirements and/or specifications of the respective
manufacturers, be sure to review the respective manufacturer's
documentation!
The following components are used on the system maintenance unit (X) and the
extension for sealing air on the glass scales (Y):
1. Switch-on valve (manual)
2. Pressure regulating valve for system pressure 6–8 bar
3. Pressure gauge for system pressure
4. Filter insert with retained particle size 40 µm with automatic condensate drain
5. Filter insert with retained particle size 5 µm with automatic condensate drain
6. System pressure monitoring
Maintenance Summary - 4.000 Operating hours
73
Maintenance Instructions
G200.2
DIE078EN - 05.06.2019