E. EMC environment guidance
Guidance and Manufacturers Declaration - Electromagnetic Emissions
The BPBPIO210/220 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BPBIO210/220 should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions CISPR 11
Group 1
The [ME EQUIPMENT or ME SYSTEM] uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions CISPR 11
CLASS A The BP210/220 is suitable for use in all establishments other
than domestic, and may be used in domestic establishments
and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes, provided the following warning is heeded:
Warning: This equipment/system is intended for use by
healthcare professionals only. This equipment/ system may
cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures,
such as re-orienting or relocating the [ME EQUIPMENT or
ME SYSTEM] or shielding the location.
Harmonic Emissions
IEC 61000-3-2
N/A
Fluctuations/Flicker
Emmisions
IEC 61000-3-3
N/A
Guidance and Manufacturers Declaration - Electromagnetic Emissions
The BPBPIO210/220 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BPBIO210/220 should assure that it is used in such an environment.
IMMUNITY Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment
Guidance.
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for Power
Supply Lines
± 1 kV for Input/
Output Lines
± 2 kV for Power
Supply Lines
± 1 kV for Input/
Output Lines
Main power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Earth
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Earth
Main power quality should be that
of a typical commercial or hospital
environment.