31
B. Regarding the Equipment
Status
Cause
Sloution
Power does not
turn on.
Check the inserted batteries.
Polarities may not be matching.
Check the battery polarities and
reinstall the batteries to match with
‘+’ and ‘-’ marks.
Batteries are exhausted.
Replace all batteries with new AA batteries.
The controller
does not work,
or it works
abnormally.
The Controller 4-Pole Connector is
not connected properly.
Plug the Controller 4-Pole Connector
tightly and push it all the way in.
The controller is damaged.
Please contact InBody Customer Center
(080-501-3939).
Pulse is not
being measured.
Blood pressure is measured too
rapidly.
Keep the decompression speed at
2 to 4 mmHg/sec.
Patient may be moving during the
blood pressure measurement.
Make sure that the patient is relaxed and
stable. Also, prevent them from talking
during the measurement.
Pulse maybe irregular or very
weak.
Impossible to measure the pulse. Please
use another device to measure the pulse
in these cases.