6. Power on Converter
After connecting according to "5. Wiring method", turn on the power and the "POWER"
lamp on the front of the converter will light up (Fig.6-1).
Memo
(1) This product takes about 30 seconds from power-on to stability.
(2) If "POWER" lamp is off even though power is being supplied, sensor cable may not
be connected properly, or sensor cable may be broken. In such a case, turn off
power supply to the converter once, check that sensor cable is connected correctly,
and then turn on power again.
VM-5011A
Power
IEPE Sensor
signal converter
Fig.6-1 POWER Lamp Indication
7. Change of Measurement Range and DC Output
Converter can change measurement range and DC output. Follow the procedure below.
Setting Method
(1) Turn off power supply to converter and disconnect all wiring connected to converter.
(2) Remove the transducer from the DIN rail.
(3) Push lock hooks on the top and bottom of converter with a flat-blade screwdriver, etc. to
release the lock, then hold upper housing and pull it out from base housing (see Fig.7-1).
Fig.7-1 Pulling out upper housing
Hold upper housing
and pull it out
When returning to original
position, align board with
guide (groove).
substrate
Unlock lock hook
using a flat-blade
screwdriver, etc.
Base Housing
Upper Housing
Status of POWER Lamp
OFF
(1) Power is not supplied
(2) Sensor is not connected
(3) Sensor cable is broken
ON
Normal (Measurement Status)
Memo
If sensor is not connected or sensor cable is
broken, output of 4-20mA will be about 2mA
(0V for 0-10V).