10
-
VM-5011A Instruction Manuals
(TVE-6-4188E / Version 1.20)
(4) Change measurement range and DC output according to Fig.7-2 below.
“JP” and “SW” Setting Item
Setting
Factory Default
JP1
E acceleration
DC Output
1-2
4-20mADC
✔
2-3
0-10VDC
JP2
Velocity
DC Output
1-2
4-20mADC
✔
2-3
0-10VDC
SW1
(*)
E acceleration
Measuring Range
1
0
~
10m/s
2
E rms
2
0
~
25m/s
2
E rms
3
0
~
50m/s
2
E rms
✔
4
0
~
100m/s
2
E rms
SW2
(*)
Velocity
Measuring Range
1
10mm/s rms
2
25mm/s rms
3
50mm/s rms
✔
4
100mm/s rms
(*) Do not turn on more than one switch.
Fig.7-2
Setting of Measurement Range and DC Output
Note
(1) Do not change jumper or adjust variable resistance other than the above.
Correct measurement may not be possible.
(2) This product is a precision device. Please handle with care.
(3) Please do not forget to record and manage the settings.
(5) After completing the settings, insert upper housing into base housing, check that lock
hook is locked, attach it to the DIN rail, and reconnection each cable.
JP1
JP2
SW1
SW2
ON
Turn on the number of
the range to be set.
(this figure is No.3)
1
Enlarged View of SW1 and SW2
2 3 4
OFF
Jumper between 1-2 (4-20mA)
or 2-3 (0-10V).
Enlarged View of JP1 and JP2
3
2
1