The following disinfection solutions have
been tested and their compatibility with
the components of the probes has been
proved:
•
Alkazyme
•
Klenzyme
•
Cidezyme
•
Nu-Cidex
•
Anioxyde
•
Gigasept-FF
•
Steranios
•
Cidex OPA
It is strongly recommended to use one of
these solutions in order to avoid any dete
-
rioration during the cleaning and disinfec
-
tion procedure.
In order to ensure proper cleaning and
disinfection, please follow the following
procedure:
1. After every patient examination, wipe
the ultrasound transmission gel off the
probe.
2. Unplug the probe connector from the
system.
3. Wipe the probe and cable with a
clean soft cloth that has been dam
-
ped in a solution of mild soap and
water.
4. Remove any cleaning solution residue
with a soft clean cloth damped in
sterile water.
5. Dry with a clean and dry cloth.
Do not rub the probe with an abrasive
sponge. Use a soft cloth or towel.
The following procedures are known to
damage transducers. They can damage
both the electrical safety features and the
acoustic performance of the probes.
Do not use the following procedures:
•
Gas sterilization.
•
Ultraviolet sterilization
•
Dry heat sterilization
•
Autoclaving
•
Soaking a transducer in a chlorine
bleach solution.
!
WARNING:
The use of any other disinfecting solution
than the ones indicated above is
dangerous as it may damage the probe
components. Check the list of specified
agents carefully