Use keys under “Next/Previous to move
along the menu.
CFM box position or size can be adjusted using Touchpad or
Cursor keys in Pre-trace CFM mode.
Probe
Labels displaying settings of the menu
Date, Time
Battery capacity
5.CFM mode and menu
F 1
!
2
@
3
?
4
+
5
%
6
-
7
&
8
/
9
(
0
)
T
AB
Q W E R T Y U I O P D
ELETE
C
APS
A S D F G H J K L E
NTER
S
HIFT
Z X C V B N M ;
T
EXT
P
ICTO
A
RROW
PATIENT
REPORT
PROTOCOL
PROBE
ZOOM
TGC
SETTINGS
CLEAR
FREQUENCY
FOCUS
M CINELOOP
PWD
MEASURE REVIEW
CFM
B
SELECT
DEPTH
GAIN
DEPTH
GAIN
FREEZE
UPDATE
STORE
CFM mode and menu.
Press Select to switch from Pre-trace
CFM to Active CFM.
Keys to change settings of the menu.
Steering of CFM box on a linear probe.
Press Update to switch from Active
CFM mode to Pre-trace CFM mode.
Power
CFM transmit Power.
Rejection of movements
from vessel walls and
tissues. High Wall Filter, slow
speeds are rejected
Wall Filter
Persistence
Frame averaging
on CFM image.
MinVelocity
Rejects the color on slow
velocities. A high Min Ve
-
locity index cuts the color
and slows velocity
Spatial Filter
Smoothening of the
CFM image.
Vessel Size
Optimization of colour
sensibility according to the
vessel size.
Invert
Inversion of the colour.
CFM scale colour is
also inverted.
Packet Size
High index increases CFM
sensitivity and decreases
CFM frame rate.
PRF
Select proper value depen
-
ding on the speed of the
measured flow
Burst
High index increases
CFM sensitivity and
decreases CFM
spatial resolution.
Labels displaying settings of the menu
in CFM mode e.g. WFillter- Wall Filter,
VesSize-Vessel Size, SFilter-Spatial Filter.
Active CFM contextual menu:
CFM gain adjustment.