G P I B / I E E E 4 8 8 . 2 R E M O T E O P E R A T I O N
Common Commands and Queries
01_10
OMM-6810B
41
C H A P T E R
3
The SERVICE REQUEST ENABLE QUERY allows the user to determine the contents of
the Service Request Enable Register.
Response
Brinary Integer Sum of enabled bits in Figure 3.11. The value must be between 0 and
255.
The STATUS BYTE QUERY allows the programmer to read the Status Byte described in
Figure 3.11.
Response
Binary Integer Sum between 0 and 255. See Figure 3.11.
The SELF-TEST QUERY initiates an internal self-test and returns a response when
complete.
Response
0 - Self test completed with no errors
1 - Self test was not completed or encountered errors
The WAIT-TO-CONTINUE command prevents the instrument from executing any further
commands until the No-Operation-Pending flag is true. This allows the programmer to
make the 6810B wait for the completion of an operation before continuing.
*SRE?
*STB?
*TST?
F
RONT
P
ANEL
R
EMOTE
*WAI
:
?
*
SRE
:
?
*
STB
:
?
*
TST
:
*
WAI