C O M M A N D R E F E R E N C E
Command Paths
01_10
OMM-6810B
89
C H A P T E R
4
The WAVE:AUTO command enables/disables the wavelength measuring feature of the
OMM-6810B.
Parameters
The <nrf value> is a boolean value of 1 or ON to enable wvelength measuremes, 0 or
OFF to disable wavelength measurements.
Notes
This command has the same effect as pressing the WAVELENGTH AUTO button to
control the wavelength measurement. If the unit is measuring wavelength the indicator
above the WAVELENGTH AUTO button is lit. If the unit is not measuring wavelength then
the WAVELENGTH AUTO indicator is not lit and the HOLD indicator is lit. When this
command is issued, the last measured wavelength remains on the display until changed
by the WAVE command.
The status of autoranging may be queried by issuing the WAVE:AUTO? query.
Example
“WAVE:AUTO ON” - action: enables wavelength measurements
WAVE:AUTO
F
RONT
P
ANEL
R
EMOTE
:
<white
space>
<nrf>
:
WAVE
AUTO