C O M M A N D R E F E R E N C E
Command Paths
70
OMM-6810B
C H A P T E R
4
The POWer:AUTO command controls the electrical gain autoranging function.
Parameters
The <nrf value> is a boolean value of 1 or ON to enable autoranging, 0 or OFF to disable
autoranging.
Note
Thsi command has the same effect as pressing the POWER AUTO button to control the
autoranging. Auto/manual ranging modes have different effects on the front panel power
display. The status of autoranging may be queried by issuing the POWer:AUTO? query.
Example
“POWER:AUTO ON” - action: enables the autoranging function.
“POW:AUTO OFF” - action: disables the autoranging function.
The POWer:AUTO? query returns the status of the autoranging mode.
Parameters
None.
Response
where the response data is a boolean value representing the status of the autoranging
mode.
0 = manual gain ranging mode
1 = auto gain ranging mode
Notes
This command has the same effect as viewing the POWER AUTO indicator on the front
panel of the multimeter. Auto/manual ranging modes have different effects on the front
panel power display. Autoranging control can be controlled by issuing the POWer:AUTO
command.
Examples
“POWER:AUTO?” - reponse: 1, means the unit is autoranging the electrical gain range.
“POW:AUTO?” - response: 0, means the unit is in manual gain ranging mode.
POWer:AUTO
F
RONT
P
ANEL
R
EMOTE
POWer:AUTO?
F
RONT
P
ANEL
R
EMOTE
:
<white
space>
<nrf>
:
POWER
AUTO
:
:
POWER
AUTO
?
<response
data>
<response
terminator>
;