C O M M A N D R E F E R E N C E
Command Paths
94
OMM-6810B
C H A P T E R
4
The WAVE:OFFset? query returns the value of the Wavelength User Calibration offset
value. This is a number that is added to wavelength measurements.
Parameters
None.
Response
where the response data is a floating point number representing the wavelength user
calibration factor.
Notes
This is a GPIB command only. The offset is entered via the WAVE:OFFset command or
by executing the user calibration as described in Chapter 6 of the optical measurement
head instruction manual. The wavelength offset is a number that is added to future
wavelength measurements as shown below:
(corrected) =
(uncorrected) + cal_factor
Examples
“WAVE:OFFSET?” - response: 0.5, depending upon the present wavelength display
mode, either 0.5 nm or 0.5 wavenumbers is added to the wavelength measurement.
“WAVE:OFF?” - response: -2.0, depending upon the present wavelength display mode,
either 2.0 nm or 2.0 wavenumbers is subtracted from the wavelength measurement.
WAVE:OFFset?