MDtest (100-140 V) - User’s Manual
6
Instrument main menu
•
IdirEq(R)
•
I60601Earth(R) + SFC, variant Open N
•
I60601Touch(R) + SFC, variant Open N or Open PE
•
I60601Pat(R) + SFC, variant Open N or Open PE
If measurement of a DC component is enabled, that measurement cycle
will take longer.
The value is transmitted to the PC software.
6.5.3
Menu - Self-test
This option can be performed to verify the functionality of the MDtest. Disconnect
equipment under test before the self-test. The self-test does not replace calibration.
The procedure is described in chapter 3.2.
6.5.4
Menu - Date of next calibration
In this menu, you can set the next calibration date. The measuring device starts
to display a warning message 45 days before the expiration date, e.g.,
Date of the
next calibration: 01.01.2018
. This information is displayed only once a day. If
the calibration has expired, there is the message, e.g.,
Calibration expired on
01.01.2018
. Enter into the menu, the date of the next calibration is performed by
pressing the
[ent]
key. The date is inserted using the keyboard device. Confirm the
selection by pressing the
[ent]
key or press the
[esc]
key to return without saving
the changes.
The ’.’ characters are automatically skipped in the editor.
6.5.5
Menu - Erase database
Use this option to erase the entire main memory device, i.e., database appliances.
The essential instrument settings and a list of test sequences remain preserved.
Erasing the memory must be confirmed by pressing the
[Yes]
key.
How long it takes to erasing entries depends on the memory usage. It
can take up to 2 minutes when all memory locations are occupied.
6.5.6
Menu - Measurement locking
This option allows you to automatically lock every measurement. When enabled,
each time measurement is displayed the symbol
AUT.ARET
. The switch activation
is done by pressing the
[ent]
key.