MDtest (100-140 V) - User’s Manual
4
Measurements
4.5
Enclosure leakage current according to EN 62353
4.5.1
Alternative method - IaltEq
Figure 8: IaltEq measurement
Outlet conditions:
- Normal
- Open PE
Do not touch exposed conductive parts of the tested appliance while
measurement is in progress –
RISK OF ELECTRIC SHOCK!
The test voltage is max. 127 V / 50 or 60 Hz, and the short circuit
current is approximately 1.5 mA.
Leakage current measurement is also possible if the MDtest is con-
nected to the IT mains.
Procedure:
•
Press the
[1 I
ALT
]
key and release it.
•
If the measurement with open PE can be performed, press the
[OpenPE]
key.
•
For a one-phase appliance with movable power supply cord, plug in its mains
plug to Test Socket
1
and plug the applied part into the
PAT
terminal. If
necessary (e.g. protective class II appliances), plug the test lead into the
PE
terminal (if Open PE is selected plug into the
PAT
terminal).
•
Switch on tested appliance’s mains switch.
•
Press the
[TEST]
key.
•
Press the test tip against the exposed conductive part of the tested appliance.
•
The last measured value is displayed after finishing the measurement.
•
Switch off the tested appliance’s mains switch.